Home Measurement Technology Leak test systems 

Leak testing by the pressure decay method

Leak testing systems

The test systems of the CoM4.Leak series measure the pressure decay of a test specimen against a leak-tight reference volume. Test specimen and reference volume are brought to the same test pressure and isolated; a differential pressure sensor records the difference between the two. Two models are available: the CoM4.Leak-Compact for test specimens with a connection and the CoM4.Leak-ClosedBox for parts that cannot be adapted.

Leak testing systems
  • ±20 mbar measuring range of the differential pressure sensor in both models
  • Capability Repeat measurements run as a program in the instrument in both models: capability analysis of the measured value in the CoM4.Leak-Compact, measurement system analysis in the CoM4.Leak-ClosedBox.
  • Ethernet interface of both models; the CoM4.Leak-ClosedBox also has USB 2.0 for updates.

The models at a glance

Test specimen with or without a connection: choose the model that suits your part.

  • Specimens with a connection

    CoM4.Leak-Compact

    The model for adapted test specimens in series production and incoming inspection

    Test pressure
    variable up to 6 bar gauge
    Resolution of pressure decay
    0.1 Pa/s
    Test programs
    up to 95
  • Test chamber for small parts

    CoM4.Leak-ClosedBox

    The part is inserted: a drawer-type test chamber for parts without a connection

    Leakage
    0.1 – 5 mL/min
    Accuracy
    < 0.5 % of reading
    Gauge pressure
    0.1 – 7 bar

Leakage measurement system

CoM4.Leak-Compact

CoM4.Leak-Compact
Test pressure
variable up to 6 bar gauge
Resolution of pressure decay
0.1 Pa/s
Test programs
up to 95

Overview

The CoM4.Leak-Compact measures the pressure decay of the test specimen against a leak-tight reference volume. The measured value is available either as standard volumetric flow or as pressure decay.

An enclosed gas volume loses pressure through leakage and through thermal equalisation. Test specimen and reference volume are filled together to the same test pressure; the supply then closes and the gas settles. For measuring, the isolation valves close, and the differential pressure sensor measures the rise of the differential pressure between reference volume and test specimen. The thermal share acts on both sides and does not enter the difference; what remains is the leakage. It becomes a quantitative leakage rate as soon as the volume of the test specimen is known – with the test leaks, the instrument determines it by itself.

Main features

  • Volume and leakage of unknown test specimens determined fully automatically with the test leaks
  • Cycle time optimised by curve set analysis
  • Self-test of internal leak-tightness and of function and leakage of the test leaks
  • Two control modes for pressure control, fast or precise, optimisable at the relevant pressure points
  • Capability analysis of the measured value from repeat measurements
  • Debug programs for fault and leak tracing on the customer-side adaptation
  • Calibration program for the built-in pressure sensors

Technical data

Version
CoM4.Leak-Compact-06bar-01
Controller
CoM4.SYS
Test pressure
variable, 0 – 6,000 mbar gauge
below 500 mbar the pressure control sometimes starts with a marked delay; this range is not suited to fast repeat testing
Differential pressure
−20 – +20 mbar
Resolution of pressure decay
0.1 Pa/s
guide value for the smallest measurable pressure decay
Sensors
PDP2 (absolute pressure): 0 – 7 bar abs., overload 14 bar, linearity deviation 0.025 % of span
PDP2 (differential pressure): −20 – +20 mbar, common pressure −1 – 16 bar, overload forward 16 bar, reverse 1 bar, linearity deviation 0.025 % of span (span 40 mbar, corresponding to 0.01 mbar)
The figure is the linearity deviation of the sensor, not the measurement uncertainty of the leak test.
Test programs
up to 95, definable as type files
Ambient conditions
10 – 40 °C, 0 – 90 % RH, non-condensing
atmospheric pressure, installation free from vibration
Ingress protection
IP 40
Medium
Compressed air: dust-free (filtered to 5 µm), oil-free, non-condensing, non-corrosive
Process connections
Test specimen: G 1/8
Test leaks 1 and 2: G 1/8 each
Compressed air supply: female thread, fitting of your choice
Enclosure
19-inch rack insert, 4 U
rack enclosure, rack enclosure with handles or desktop enclosure with feet and handles
Dimensions (W × H × D)
450 × 190 × 550 mm
height of the desktop enclosure with feet
Weight
15 kg
Power supply
100 – 240 V AC
Interfaces
Ethernet: COMM, TetraTec Net-IO
Operating language
English and German
Directives, marking
EMC Directive, Low Voltage Directive; CE
Certificates supplied
DAkkS calibration certificates of the pressure sensors
DAkkS calibration certificates of the test leaks, if ordered
recalibration either as DAkkS or as factory calibration
Option
Capillary test leaks for volume determination and the self-tests – strongly recommended, both functions require them

Downloads

Your request

Select what applies – your selection is then entered in your message in the contact form.

  1. Enclosure

Leakage measurement system

CoM4.Leak-ClosedBox

CoM4.Leak-ClosedBox
Leakage
0.1 – 5 mL/min
Accuracy
< 0.5 % of reading
Gauge pressure
0.1 – 7 bar

Overview

The CoM4.Leak-ClosedBox tests small parts that cannot be adapted, tubed or opened. They are placed in a closed test chamber, tested and removed again; the sequence runs fully automatically.

The same test pressure is set electronically in the chamber and in a leak-tight reference volume, after which the isolation valves close. If the part leaks, the differential pressure sensor measures the pressure decay relative to the reference volume. Since the pressure decay depends on the test volume, the system converts it into a leakage rate by calibration with built-in test leaks.

Main features

  • Chamber insert to suit your part geometry, made to drawing
  • Self-calibration via built-in test leaks
  • Filling, settling and measuring adjustable for each test task, with variable shock filling

Technical data

Controller
CoM4.SYS software on Beckhoff system
Leakage
0.1 – 5 mL/min
accuracy < 0.5 % of reading, measurement uncertainty (k = 2) < 1.0 % of reading
Gauge pressure
0.1 – 7 bar, 0.01 % FS
Temperature
0 – 50 °C, 0.15 °C
Calibrated ranges
apply to air
Sensors
Gauge pressure sensor 0: 0 – 16 bar, overload 3 × full scale
Gauge pressure sensor 1: 0 – 10 bar, overload 12 bar
Differential pressure sensor: −20 – +20 mbar, overload 2 × full scale
Pt100 temperature sensor: 0 – 75 °C, max. 75 °C
Operating conditions
Inlet pressure max. 10 bar
Inlet temperature 0 – 50 °C
Humidity 0 – 100 %, non-condensing
Ambient conditions
0 – 40 °C, 0 – 100 %, non-condensing, atmospheric pressure
Ingress protection
IP 20
Medium
clean, dry, non-condensing air
purity to ISO 8573-1; 5 µm filter and oil/water separator required
Part holder
drawer with moulded insert
Enclosure
19-inch desktop enclosure, 10 U, without handles
495.5 × 449 × 444 mm
Weight
28 kg
Display
external monitor with Beckhoff HMI
Power supply
90 – 260 V AC, 50/60 Hz, max. 80 W, IEC connector
Interfaces
Ethernet, USB 2.0 (type A, for updates)
Standards and directives
EN 61010-1
Options
Desktop or rack enclosure, also with handles
Integration of the mechanics into the production line

Downloads

Your request

Select what applies – your selection is then entered in your message in the contact form.

  1. Installation

Around leak testing

Capillary test leaks monitor leak testing systems or simulate leaks during set-up. For tasks with their own mechanics and handling we build complete test rigs.